Los Alamos National Laboratory

Environment » Air » NEWNET

Tonopah
Tonopan, NV

View different station

EnteredByCommentView 
3/11/1999Ken GilesGamma data on march 2,1999 of 116uR and on march 10,1999 of 116uR were from source checks preformed by EPA personal doing routine maintance on the equipment.Station InfoData Graph
2/24/1999 on 23 febuary,1999 the PIC sensor was replaced with cal. unit #L-3104.Station InfoData Graph
8/21/1998Larry SandersThe station was loaded with NEWNET version R35 software on 8/21/98 at 15:55Z. This software provides improved accuracy for radiation measurements in the low range. The station was re-wired with a door sensor and new voltage divider for the radiation sensor. A new 12 volt battery was be installed. The PIC 300volt battery voltage was not replaced, the battery was reading 292volts. Currently, this station only has a radiation sensor. EPA will install meteorological sensors later. All meteorological inputs will remain shorted until the meteorological sensors are installed.Station InfoData Graph

This site passed IRM-CAS quality check